Error codes – IKA RV 10 digital FLEX User Manual
Page 34

34
Any malfunctions during operation will be identified by an error message on the display (only RV 10 digital).
Once a serious error message has been displayed, the lift moves to the top end position and the device can no longer be operated. The lift can be
operated again.
Proceed as follows in such cases:
GSwitch off device using the main switch
Carry out corrective measures
Restart device
Error Effect
Cause
Correction
code
E01
No or too little rotation
Cable break
Reduce the volume in the evaporator piston
Load too large
Motor blocked
E02
No communication with the
Interface hidden or dirty
Check / clean the interface
heating bath
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- contact the IKA
®
service department,
- send the device for repair, including a short description of the fault.
RV 10.2029
NS 24/40 Distillation spider with 6 sleeves
RV 10.2030
NS 24/40 Distillation spider with 12 sleeves
RV 10.2031
NS 24/40 Distillation spider with 20 sleeves
RV 10.610
Distilling sleeve 20 ml
RV 10.606
NS 29/32 Distillation spider with 5 flasks 50 ml
RV 10.607
NS 29/32 Distillation spider with 5 flasks 100 ml
RV 10.608
NS 24/29 Distillation spider with 5 flasks 50 ml
RV 10.609
NS 24/29 Distillation spider with 5 flasks 100 ml
RV 10.2032
NS 29/42 Distillation spider with 5 flasks 50 ml
RV 10.2033
NS 29/42 Distillation spider with 5 flasks 100 ml
RV 10.2034
NS 24/40 Distillation spider with 5 flasks 50 ml
RV 10.2035
NS 24/40 Distillation spider with 5 flasks 100 ml
RV 10.3000
Extension plate
RV 10.8001
Seal, PTFE compound
PC 1.1
Cable
Labworldsoft
®
Error codes