REO basic
RET basic safety control
Base plate
Dimensions
mm
ø 135
Material
Stainless steel
Heating
Heating power (-5%/+10%) at 230 V
W
600
Heating power (-5%/+10%) at 115 V
W
600
Heating power (-5%/+10%) at 100 V
W
450
Adjustment and display resolution
K
1
Surface temperature
°C
Ambient temperature - 340
Temperature sensor PT 1000 variation
K
< + (0,15 +0,002xITI)
DIN EN 60751 Kl. A
Temperature variation max. at 100 °C
K
+1,5
Heating plate temperature variation
K
+5
Heating plate control hysteresis,
K
+5
no container, center of heating plate at 100 °C
Control hysteresis,
K
+1
with temperature sensor PT 1000*
with ETS-D5*
°C
+0,5
with ETS-D6*
°C
+0,2
Adjustable safety circuit
Safety temperature limit (adjustable)
°C
50 - 360 (+10)
*
The control accuracy values given were determined using the following set-up: 500 ml water in 600 ml glass beaker, magnetic stirring bar 40, 600 rpm, 50 °C
Subject to technical changes!
23
092008
In accordance with IKA warranty conditions, the warranty period
is 24 months. For claims under the warranty please contact your
local dealer. You may also send the machine direct to our factory,
enclosing the delivery invoice and giving reasons for the claim.
You will be liable for freight costs.
Warranty