8

9

Source language: German

EN

Page

CE-Declaration of conformity

8

Explication of warning symbols

8

Safety instructions

9

Unpacking 10
Correct use

10

Commissioning 10
Correction of malfunctions

11

Maintenance and cleaning

11

Accessories 12
Technical data

12

Warranty 12

Contents

We declare under our sole responsibility that this product is in compliance with the regulations 2006/95/EG and
2004/108/EG and conforms to the standards or standardized documents DIN EN IEC 61010-1, -2-051 und DIN EN IEC
61326-1.

DANGER

CAUTION

WARNING

General hazard

This symbol identifies information that is of absolute importance to ensu-

re health and safety. Failure to observe this information may be detrimental

to health or may result in injuries.

This symbol indicates information which is important for ensuring that

the appliance functions without any technical problems. Failure to ob-

serve this information could damage the appliance.

This symbol indicates information which is important for proper use of

the appliance and/or ensuring that the appliance functions correctly.

Failure to observe this information can lead to inaccurate results.

Attention - Risk of damage due to magnetism.

CE-Declaration of conformity

Explication of warning symbols

ATTENTION

Safety instructions

• Only process media that will not react dange-

rously to the extra energy produced through
processing. This also applies to any extra ener-
gy produced in other ways, e.g. through light
irradiation.

• Please observe the operating instructions for

any accessories used.

• Safe operation is only guaranteed with the ac-

cessories described in the ”Accessories” chap-
ter.

• The centre of gravity of the assembly must lie

within the surface on which it is set up.

• The appliance can only be disconnected from

the mains supply by pulling out the mains plug
or the connector plug.

• The socket for the mains cord must be easily

accessible.

• The device will automatically restart following

any interruption to the power supply.

• It may be possible for wear debris from rota-

ting accessory parts to reach the material being
processed.

• When using PTFE-coated magnetic bars, the

following hase to be noted: Chemical reactions
of PTFE occur in contact with molten or solute
alkali metals and alkaline earth metals, as well
as with fine powders of metals in groups 2
and 3 of the periodic system at temperatures
above 300 °C - 400 °C. Only elementary fluo-
rine, chlorotrifluoride and alkali metals attack
it; halogenated hydrocarbons have a reversible
swelling effect.

(Source: Römpps Chemie-Lexikon and ”Ulmann”,
Volume 19

)

For protection of the equipment

• The socket used must be earthed (fitted with

earth contact)

• The appliance may only be opened by experts.
• The voltage stated on the type plate must cor-

respond to the mains voltage.

• Do not cover the device, even partially e.g. with

metallic plates or film. This results in overhea-
ting.

• Protect the appliance and accessories from

bumps and impacts.

• Ensure that the base plate is kept clean.

For your protection

• Read the operating instructions in full before

starting up and follow the safety instructions.

• Keep the operating instructions in a place whe-

re they can be accessed by everyone.

• Ensure that only trained staff work with the ap-

pliance.

• Follow the safety instructions, guidelines, occu-

pational health and safety and accident preven-
tion regulations.

• Caution - Magnetism! Effects of the magnetic

field have to be taken into account (e.g. data
storage media, cardiac pacemakers ...).

• Wear your personal protective equipment in ac-

cordance with the hazard category of the me-
dia to be processed. Otherwise there is a risk
from:
- splashing and evaporation of liquids
- ejection of parts
- release of toxic or combustable gases.

• Set up the appliance in a spacious area on an

even, stable, clean, non-slip, dry and fireproof
surface.

• The feet of the appliance must be clean and

undamaged.

• Check the appliance and accessories before-

hand for damage each time you use them. Do
not use damaged components.

• Reduce the speed if

- the medium splashes out of the vessel be-

cause the speed is too high

- the appliance is not running smoothly
- the container moves on the base plate.

• Beware of hazards due to:

- flammable materials
- combustible media with a low boiling tem-

perature

- incorrect container size
- overfilling of media
- unsafe condition of container.

• The appliance may heat up when in use.
• The base plate can heat up due to the action of

the drive magnets at high motor speeds, even
if the heater is not operational.

• Process pathogenic materials only in closed

vessels under a suitable extractor hood. Please
contact IKA if you have any questions.

• Do not operate the appliance in explosive at-

mospheres, with hazardous substances or un-
der water.

ATTENTION