GENERAL:
1. All service work should be performed in a clean area.
2. Care should be taken to avoid scratching any sealing surfaces or allowing metal chips to come in contact with
the Driven Magnet Assembly.
3. An even, light coat of high vacuum silicone grease (Dow Corning or equiv.) should be applied to all seals or O-
Rings.
4. O-Rings with chamfers should always be installed with the chamfer up, away from the groove.
5. All assembly screws should be tightened evenly and in an alternating pattern.
6. Only Micropump factory authorized replacement parts should be used when servicing any Micropump products.
SPECIFIC:
1. Do not attempt to remove Driving Gear Shaft (2) from Pump Body (1). The Shaft (2) position is preset at the
factory. (If replacement of shaft is necessary, pump must be returned to the factory for repair.)
2. Install the Driving Gear (7) and Driven Gear (4) over the Shafts (2 & 3)
3. Install the Suction Shoe (9) and install the Suction Shoe Spring (10) per the spring setting procedure below.
4. Engage the Driven Magnet (8) with Driving Gear (7). Gears and Magnet should rotate freely.
5. Complete assembly as shown above.
SERVICE INSTRUCTIONS: L23814
MODEL 180 - "M" & J MOUNT