Troubleshooting, Storage and shipping – BNC 1621M User Manual
Page 22

4. TROUBLESHOOTING
The list of possible problems and their solutions are specifi ed in Table 4.1
5. STORAGE AND SHIPPING
5.1. Packaged devices may be shipped by any method of closed transport
at the air temperature from -50 to +50°C .
5.1. Packaged devices may be shipped by any method of closed transport
5.1. Packaged devices may be shipped by any method of closed transport
5.2. Packaged devices should be secured during transport.
5.3. When carried by sea, packaged devices should be sealed in
plastic bags with silicagel.
5.4. When carried by air, packaged devices should be placed
in sealed compartments.
5.5. Devices should be stored in the manufacturer’s package at an air
temperature ranging from -15 to +50°C and relative
5.5. Devices should be stored in the manufacturer’s package at an air
5.5. Devices should be stored in the manufacturer’s package at an air
humidity up to 95% at a temperature of 35°C.
temperature ranging from -15 to +50°C and relative
temperature ranging from -15 to +50°C and relative
5.6. Devices without package should be stored at an air temperature from
10 to 35°C and relative humidity up to 80% at a temperature of 25°C.
5.6. Devices without package should be stored at an air temperature from
5.6. Devices without package should be stored at an air temperature from
5.7. The storage location should be free of dust, chemical vapors,
aggressive gases and other substances that may cause corrosion.
Note: If a defective battery is used the Er1-Er7 message may appear.
Attention! When the Er1-Er7 message appears, press any button. When the error message appears for the
A
A
second time (approx. 15 min) the dosimeter is not to be used.
Problem
The LCD indicates “bAt”
message
No indictions on LCD
The device does not respond
to pressing a button. the LCD
indicates incorrect symbols
LCD indicates Er1-Er7
Possible Cause
Battery discharge
Battery discharge
Battery inserted incorrectly
Microprocessor error condition
Dosimeter failure
Solution
Replace Battery
Replace Battery
Insert battery correctly
Remove battery and
insert again after 5 minutes
Return instrument for repair.
21