beautypg.com

Specifications – B&K Precision 1735A - Manual User Manual

Page 7

background image

7

SPECIFICATIONS






























OUTPUT VOLTAGE:

0 to 30VDC, coarse and fine

adjustment.

OUTPUT CURRENT:

0 to 3A, coarse and fine adjustment.


CONSTANT VOLTAGE OPERATION

Voltage Regulation
Line (108-132 V):

0.01% + 3mV.

Load (no load to full load):

0.01% + 3mV.


Recovery Time:

100 µs typical.


Ripple Voltage
Peak-to-Peak:

2mV, typical

RMS:

1mV.


Temperature Coefficient

(0° to +35 °C):

300 PPm/°C


CONSTANT CURRENT OPERATION

Adjustable Current Limits:

5% to 100%

Current Regulation

Line (108-132 V):

0.2% + 3mA

Load

0.2% + 3mA


Current Ripple:

3mA typical.


METERING

Voltmeter

4 digit Green LED Display

Range:

0 to 99.99 V.

Accuracy:

(0.5% + 9 digits)*.


Ammeter

4 digit Red LED Display

Range:

0 to 9.999A.

Accuracy:

(0.5% + 9 digits)*.

(* = see note 1 at end of specifications)

POWER REQUIREMENTS:

Domestic: 120 VAC ±10%, 60 Hz
International: 120/220/230/240 VAC - 10%, 50/60 Hz

POWER CONSUMPTION:

Approximately 180W or less at
full load.


PROTECTION:

Reverse polarity protection,

Current limiting.