Accuracy
The accuracy of the Ascensia
C
ONTOUR
Blood Glucose Monitor-
ing System has been assessed in a correlation study conducted
by healthcare professionals. In this study, 108 fresh capillary
blood specimens were tested with the Ascensia
C
ONTOUR
Blood
Glucose Monitoring System by 108 untrained people with dia-
betes using three test strip lots. The reference measurement
was done on the YSI Glucose Analyzer. Ascensia
C
ONTOUR
data
compare well with the results from the YSI method.
Regression Statistics (Plasma/Serum Reference):
plasma reference glucose range: 4.2 to 22.0 mmol/L
Precision
A study was conducted with the Ascensia
C
ONTOUR
Blood
Glucose Monitoring System with fi ve heparinized whole
blood specimens with glucose levels from 2.2 to 18.0
mmol/L. Multiple replicates (n=100) were tested using mul-
tiple Ascensia
C
ONTOUR
Blood Glucose Meters and one lot
of Ascensia
M
ICROFILL
Blood Glucose Test Strips. The fol-
lowing precision results were obtained:
Plasma/Serum:
Technical Information:
Technical Information
Lot
y =
Correlation
Coeffi cient
(r)
1
1.02x – 0.01 mmol/L
0.972
2
1.03x – 0.24 mmol/L
0.963
3
0.95x – 0.02 mmol/L
0.974
Mean
SD
Coeffi cient
of
variation
2.1 mmol/L
0.10 mmol/L
4.8%
5.6 mmol/L
0.28 mmol/L
5.1%
6.5 mmol/L
0.24 mmol/L
3.7%
11.4 mmol/L
0.38 mmol/L
3.3%
18.1 mmol/L
0.65 mmol/L
3.6%