To calibrate a display manually, When done, click button set point 1 of 20, Click ok to finish – Barco DICOM THEATER LITE MGP D5 User Manual
Page 92: To reset to the latest calibration

Using TheaterWatch
MGP D5 Dicom Theater Lite
93
To calibrate a display manually:
1. Follow step 2 and 3 of the calibration with Barco LCD sensor
2. In the Calibration dialog, check the option Visually and click
the OK button.
The DICOM Optimizer dialog appears.
Figure 36: DICOM Optimizer dialog
3. Adjust the slider until there is almost no visible difference
between the background of the image above the slider and
the bitmap inside this image.
4. When done, click button Set point 1 of 20.
5. Repeat this procedure until the button is grayed. This
indicates the last point is set.
6. Click OK to finish.
To reset to the latest calibration:
1. Follow step 2 and 3 of the calibration with Barco LCD sensor
2. In the Calibration dialog, check the option Use the latest
calibration.