Air conditioning and environmental requirements, Temperature, humidity, and altitude, Air conditioning and environmental requirements 2 – Agilent Technologies 5975 User Manual
Page 28: Temperature, humidity, and altitude 2, Air conditioning and environmental requirements 28
28
Site Preparation Guide
Site Preparation
Air Conditioning and Environmental Requirements
Air conditioning considerations include temperature, humidity, airborne dust,
and exhaust venting. Each of these is considered in more detail in the
following material.
Temperature, humidity, and altitude
The 5975 Series MSD is specified for operation under the following conditions:
• Operation requires constant temperature
(variations <2
°C/h)
• Operation and storage require a noncondensing, noncorrosive atmosphere
• The temperature and humidity limits in Table 5
• Maximum altitude for operation: 2300 m (7500 ft)
Environmental control systems must maintain these temperature and
humidity ranges.
The MSD is rated for 1200 Watts (4000 BTU/h). The GC and the data system
also contribute significantly to the cooling load although the exact amounts
depend on their configurations. Additional allowances should be made for
other heat sources such as heat from other equipment, heat from adjacent
rooms, and heat from laboratory personnel.
Table 5
Temperature and humidity limits
Temperature
Humidity (relative)
Operation
15 °C to 35 °C (59 °F to 95 °F)
40% to 80%
Storage
–20 °C to 70 °C (–4 °F to 158 °F)
0% to 95%
5975 Series Site Prep.book Page 28 Thursday, January 31, 2008 9:47 AM