Power requirements

Table 4 lists the power requirements for the 5975 Series MSD and related
equipment. Extra power capacity for future additions is recommended.

Each product listed requires a dedicated circuit. The gauge controller is
powered by the MSD. The MSD, GC, and data system must each have a
separate circuit breaker. All of the equipment must share a common ground.

Power must meet the stability specifications listed in Table 4. Use a line
monitor to check power stability. If your line power is unstable, you may need
to install a line conditioner.

Table 4

Power requirements

Product

Line voltage

Maximum continuous
power consumption

Supply
circuit rating

Outlets
required

5975 Series MSD

*

120-127 VAC 60 Hz

900 VA (400 VA for

foreline pump only)

15 A

1

200-240 VAC, 50/60 Hz

900 VA

15 A

1

7890A and 6890 GCs
(standard oven) and ALS

Americas: 120 VAC

single phase

(-10%/+5%), 48-66 Hz

2250 VA

20 A

1

220/230/240V single/split phase
(-10%/+5%), 48-66 Hz

2250 VA

15 A

1

7890A and 6890 GCs (fast
oven) and ALS

Japan: 200V split phase (-10%/+5%),
48-66 Hz

2950 VA

15 A

1

220/230/240 VAC

single/split phase

(-10%/+5%), 48-66 Hz

2950 VA

15 A

1

6850 GC (standard
oven)

**

America: 120 VAC, 48-63 Hz

1440 VA

15 A

1

Japan: 100 VAC, 48-63 Hz

1440 VA

15 A

1

230 VAC, 48-63 Hz

2000 VA

8-12 A

1

6850 GC (fast oven)**

America: 120 VAC, 48-63 Hz

2400 VA

20 A

1

Japan: 200 VAC, 48-63 Hz

2400 VA

12 A

1

230 VAC, 48-63 Hz

2400 VA

11 A

1

5975 Series Site Prep.book Page 25 Thursday, January 31, 2008 9:47 AM