Power requirements, Power requirements 25 – Agilent Technologies 5975 User Manual
Page 25
![background image](/manuals/47990/25/background.png)
Site Preparation
Site Preparation Guide
25
Power requirements
Table 4 lists the power requirements for the 5975 Series MSD and related
equipment. Extra power capacity for future additions is recommended.
Each product listed requires a dedicated circuit. The gauge controller is
powered by the MSD. The MSD, GC, and data system must each have a
separate circuit breaker. All of the equipment must share a common ground.
Power must meet the stability specifications listed in Table 4. Use a line
monitor to check power stability. If your line power is unstable, you may need
to install a line conditioner.
Table 4
Power requirements
Product
Line voltage
Maximum continuous
power consumption
Supply
circuit rating
Outlets
required
5975 Series MSD
*
120-127 VAC 60 Hz
900 VA (400 VA for
foreline pump only)
15 A
1
200-240 VAC, 50/60 Hz
900 VA
15 A
1
7890A and 6890 GCs
(standard oven) and ALS
Americas: 120 VAC
†
single phase
(-10%/+5%), 48-66 Hz
2250 VA
20 A
1
220/230/240V single/split phase
(-10%/+5%), 48-66 Hz
2250 VA
15 A
1
7890A and 6890 GCs (fast
oven) and ALS
Japan: 200V split phase (-10%/+5%),
48-66 Hz
2950 VA
15 A
1
220/230/240 VAC
‡
single/split phase
(-10%/+5%), 48-66 Hz
2950 VA
15 A
1
6850 GC (standard
oven)
**
America: 120 VAC, 48-63 Hz
1440 VA
15 A
1
Japan: 100 VAC, 48-63 Hz
1440 VA
15 A
1
230 VAC, 48-63 Hz
2000 VA
8-12 A
1
6850 GC (fast oven)**
America: 120 VAC, 48-63 Hz
2400 VA
20 A
1
Japan: 200 VAC, 48-63 Hz
2400 VA
12 A
1
230 VAC, 48-63 Hz
2400 VA
11 A
1
5975 Series Site Prep.book Page 25 Thursday, January 31, 2008 9:47 AM