131

English

Blood pressure classification

SYS (mm HG)

DIA (mm HG)

Optimal

<120

<80

Normal

120 - 129

80 - 84

high - normal

130 - 139

85 – 89

Hypertension level 1

140 - 159

90 – 99

Hypertension level 2

160 - 179

100 – 109

Hypertension level 3

≥180

≥110

Definition and classification of the blood pressure values according to WHO/ISH

10. TROUBLESHOOTING (1)

PROBLEM

POSSIBLE CAUSE

SOLUTION

LCD Display shows
abnormal result

The cuff position was not correct or it
was not properly tightened

Apply the cuff correctly and try again

Body posture was not correct during
testing

Review the section “BODY POSTURE
DURING MEASUREMENT” of the instruc-
tions and try again.

Speaking, arm or body movement,
angry, excited or nervous during test-
ing

Re-test when calm and without speaking
or moving during the test