Recora COS Chair Occupancy Sensors User Manual
Recora Sensors
Best Practices
for the use of
Bed Occupancy Sensors (BOS) and
Chair Occupancy Sensors (COS)
To clean the sensors wipe with a clean, dry cloth. Avoid holding any sensor by its
connection wire while wiping.
Test for correct activation of the BOS or COS using a resistance meter like a FLUKE
Model 175. When not activated, the resistance should be greater than 10 Mohms, when
activated, the resistance should be less than 10 ohms.
In the pictured test the resistance meter is reading beyond its capability to record a
resistance reading. Interpret this as much greater than 10 Mohms.