Each adapter is specified as a single delay/thru
standard and up to seven standards numbers can
be specified into the adapter class.

Standard Class labels
Standard Class labels are entered to facilitate
menu-driven calibration. A label can be any user-
selected term which best describes the device or
class of devices that the operator should connect.
Predefined labels exist for each class. These labels
are

S

11

A, S

11

B, S

11

C, S

22

A, S

22

B, S

22

C, FWD TRANS,

FWD MATCH, REV TRANS, REV MATCH,
RESPONSE, FWD ISOLATION, REV ISOLATION,
THRU, REFLECT, LINE, and ADAPTER.

The class labels for the WR-62 waveguide calibra-
tion kit are as follows; S

11

A and S

22

A–PSHORT1;

S

11

B and S

22

B–PSHORT2; S

11

C and S

22

C–PLOAD;

FWD TRANS, FWD MATCH, REV TRANS and REV
MATCH–PTHRU; and RESPONSE–RESPONSE.

TRL options
When performing a TRL 2-PORT calibration, cer-
tain options may be selected. CAL Z is used to
specify whether skin-effect-related impedance vari-
ation is to be used or not. Skin effect in lossy
transmission line standards will cause a frequency-
dependent variation in impedance. This variation
can be compensated when the skin loss (offset
loss) and the mechanically derived impedance
(Offset Z

0

) are specified and CAL Z

0

: SYSTEM Z

0

selected. CAL Z

0

: LINE Z

0

specifies that the imped-

ance of the line is equal to the Offset Z

0

at all

frequencies.

The phase reference can be specified by the Thru
or Reflect during the TRL 2-PORT calibration. SET
REF: THRU corresponds to a reference plane set by
Thru standard (or the ratio of the physical lengths
of the Thru and Line) and SET REF: REFLECT cor-
responds to the Reflect standard.

LOWBAND FREQUENCY is used to select the mini-
mum frequency for coaxial TRL calibrations. Below
this frequency (typically 2 to 3 GHz) full 2-port
calibrations are used.

Note
The resultant calibration is a single cal set
combining the TRL and conventional full 2-port
calibrations. For best results, use TRM calibration
to cover frequencies below TRL cut-off frequency.

Calibration kit label
A calibration kit label is selected to describe the
connector type of the devices to be measured. If a
new label is not generated, the calibration kit label
for the kit previously contained in that calibration
kit register (CAL 1 or CAL 2) will remain. The pre-
defined labels for the two calibration kit registers
are:

Calibration kit 1

Cal 1

Agilent 85050B

7-mm B.1

Calibration kit 2

Cal 2

Agilent 85052B

3.5-mm B.1