"E-4" Result

What It Means:
Your blood glucose may be too high to

be read by the system, or there may be

a problem with the test strip.
What to Do:
Monitor your blood glucose again with

a new test strip. If E-4 shows on the

display window again, contact your

healthcare professional immediately.

Important:

It is recommended that you repeat the blood glucose

test with a new test strip when:

LO shows on the display window.

You obtain a low blood glucose result, but you do not have

symptoms of low blood glucose.

Your monitor displays a blood glucose result that is unusually

low, lower than 50 mg/dL.

HI shows on the display window.

You obtain a high blood glucose result, but you do not have

symptoms of high blood glucose.

Your monitor displays a result that is unusually high, higher

than 300 mg/dL.

You question your result.

Important:

A result that is incorrect may have a serious medical outcome.

Consult your healthcare professional before changing your

diabetes medication program.