beautypg.com

Zeroing, Weighings in two ranges – RADWAG WPT/P 60 Postal Scales for Packages User Manual

Page 18

background image

• Using

and

set the tare value,

• Press ,
• Program returns to weighings mode. The inscribed tare

value can be seen on the display with „–” sign,

• Tare can be inscribed anytime in weighings mode.


Notice:
You cannot inscribe a new tare value when the tare value in memory
is greater than zero. In the case of trying this the message
will be displayed and short audible signal will be emitted.



11.3. Zeroing

To ZERO the scale press:

.

The scale will display zero and following pictograms:

and

.

Zeroing is only possible within the scope of ±2% of full scale.
While zeroing outside the scope of ±2% you will see Err2.
Zeroing is possible only in stable state.


Notice:
Zeroing is possible only within the ±2% interval of the maximal range.
If zeroing is performed beyond this range the message and
short audible signal will be emitted.



11.4. Weighings in two ranges

Switching between the I range and the II range happens automatically
(exceeding Max of the I range). Weighings in the second range is
signalled by a pictogram in the top left corner of the display.

Then weighings is done with the accuracy of the II range to the moment
of returning to zero (autozero range

) where the scale switches back

to the I range.

- 18 -

This manual is related to the following products: