beautypg.com

Operator’s manual for the pressure injection cell – Next Advance PC77-MAG User Manual

Page 7

background image

5

BE SAFE. ALWAYS MAKE SURE THAT THERE IS NO PRESSURE IN THE CELL BEFORE LOOSENING THE

BOLTS, AND THAT THE BOLTS ARE SECURELY FASTENED BEFORE APPLYING PRESSURE TO THE CELL.

USERS MUST ALWAYS WEAR PROTECTIVE APPAREL, INCLUDING EYEWEAR AND MUST BE TRAINED IN

WORKING WITH HIGH PRESSURE.

Operator’s Manual for the Pressure Injection Cell

TROUBLESHOOTING

In addition to the tips given below, a

thorough list of troubleshooting tips is at

http://www.nextadvance.com/api/index.cfm/c

ms.page/i/5063/pressure-injection-cell-troubl

eshooting.

If gas leaks through the top-base

interface of the pressure injection cell, the

three bolts might not be fully tightened or the

O-ring might be nicked or dirty or missing. If

the O-ring is dirty, wash it off with a soap and

water and return it to its groove. Contact us

for a replacement O-ring.

If gas leaks around the capillary, the

ferrule should be replaced.

If you do not see sample traveling

through the capillary after a few minutes,

check that the gas tank is supplying the

desired pressure (typically on the order of

several hundred psi) and that the 3-way valve

knob is pointing toward the pressurized gas

line.

SUPPORT

A thorough list of FAQs is at

http://www.nextadvance.com/api/index.cfm/p

roducts.info/c/427/Pressure-Injection-Cell-Ca

pillary-Loader#/FAQ.

If you cannot find a good answer there,

please contact customer service by email at

[email protected]

or by telephone.

SPECIFICATIONS

Max. Pressure: 17 MPa (2500 psi)
Size: 18 cm (6.5 in.) long x 10 cm (4 in.)

wide x 19 cm (8 in.) high
Weight: 1.6 kg (3.5 lbs.)
Relative Humidity: 5 - 90% non-condensing
Operating Temperature: 4 - 40°C
Altitude: <2000m
Storage Temperature: -40 to 50°C
Meets requirements
Power Requirement (PC77-MAG model):

24V DC, 0.8 A

WARRANTY

Next Advance warrants its Products against

defects in materials and workmanship for time

periods which vary according to the Product.

Within these time periods, Next Advance will

replace or repair, without charge to the original

purchaser, any part which is defective.
Pressure Injection Cells Two years
The warranty is void if the Product is

defective due to product accident, product

modification, exposure to radiation other than

for sterilization, connection to an improper

electrical supply, lack of proper maintenance,

contamination, improper installation or misuse.

If the product is used in a manner not specified

by the manufacturer, the protection provided by

the equipment may be impaired. The warranty

shall also not apply to defects arising from fire,

flood, lightning or other conditions unrelated to

correct operation of the Product.
Next Advance’s liability is limited, at the

company’s election, to (1) refund of the

original purchaser’s purchase price for the

Product (2) repair of the Product, or (3)

replacement of the Product or defective parts.

Evidence of purchase by the original purchaser

is required. Next Advance may also request

This manual is related to the following products: