beautypg.com

Operator’s manual for the pressure injection cell – Next Advance PC77-MAG User Manual

Page 6

background image

4

BE SAFE. ALWAYS MAKE SURE THAT THERE IS NO PRESSURE IN THE CELL BEFORE LOOSENING THE

BOLTS, AND THAT THE BOLTS ARE SECURELY FASTENED BEFORE APPLYING PRESSURE TO THE CELL.

USERS MUST ALWAYS WEAR PROTECTIVE APPAREL, INCLUDING EYEWEAR AND MUST BE TRAINED IN

WORKING WITH HIGH PRESSURE.

Operator’s Manual for the Pressure Injection Cell

desired pressure.

When the injection is finished, close the

gas tank valve. Then turn the knob on the

3-way valve so that it points toward the vent.

This depressurizes the injection cell. Now

you can loosen the nut over the ferrule and

remove the capillary, loosen the three bolts,

remove the top, and remove your sample

tube.

Use a wrench to tighten the nut, forcing the ferrule

to squeeze tightly around the capillary.

USE OF ADAPTER FOR 1 ML TUBES

We supply one adapter with the pressure

injection cell so that 1 ml tubes are as easy to

work with as taller tubes.

Photograph of a 1 ml tube with a hinged cap (left)

and of a 1 ml tube with the hinge cut off and the

bottom placed in an adapter composed of soft

tubing (right).

The central hole in the base of the pressure

injection cell is sized so that 1.5 or 2 ml

sample tubes will protrude above the surface,

making placement and retrieval easy. A

(surprisingly) simple adapter will enable 1 ml

microcentrifuge tubes to protrude, too. The

adapter is a piece of soft tubing (1/4 inch

inside diameter, 3/8 inch outside diameter).

As with any tube with a hinged cap, first cut

off the hinge so that the top of the tube will fit

in the recess in the top of the pressure

injection cell. Next, slide the adapter onto the

bottom of the 1 ml sample tube, as shown in

the figure below.

STIRPLATE OPERATION

(PC77-MAG model only)

Operation of the stirplate is straightforward.

The power supply should be plugged in to the

jack at the rear. The speed control knob is on

the left. Clockwise rotation increases the

speed. The power switch is on the right.

MAINTENANCE

Periodically, the Teflon ferrule should be

replaced. Simply remove the nut covering the

ferrule, remove the ferrule, place a new

ferrule, tapered side down into the fitting, and

replace the nut. Your pressure injection cell

comes with 10 ferrules (plus one installed in

the top). Additional ferrules can be purchased

directly from Chromatography Research

Supplies (part number 214204, 1/8” to 0.4mm

PTFE ferrules), at www.chromres.com

DISCONNECTING AND

RECONNECTING THE UNIT

Before you disconnect the pressure injection

cell, mark lines on the nut and the valve,

close the valves by the pressure source, and

release any pressure inside the cell by

opening the 3-way valve. When you recon-

nect the unit, you can use the same ferrule

and tubing. Simply tighten the nut slightly

more than last time. You should feel the

wrench get very hard to turn once you go

beyond the nut’s previous position.

This manual is related to the following products: