WIKA operating instructions models 53, 54, 55

WARNING!

Before storing the instrument (following operation),

remove any residual media. This is of particular

importance if the medium is hazardous to health, e.g.

caustic, toxic, carcinogenic, radioactive, etc.

The use of liquid damping is always recommended for

temperatures near the dew point (±1 °C around 0 °C).

6. Commissioning, operation

When screwing the gauges in, the force required for this must not

be applied through the case or terminal box, rather only through the

spanner flats provided for this purpose (using a suitable tool).

Installation with

spanner

When mounting a bimetal thermometer that can be rotated and

swivelled, the specific instructions must be followed. In order to set

the indicator to the desired position, the following steps must be

taken:

1. The lock nut or union nut must be loosened at the process

connection.

2. The hexagon bolts and slotted screws at the swivel joint must be

loosened.