6. Indicator check / 7. Indicator correction

11484498.02 06/2010 GB/D/F/E

WIKA operating instructions bimetal thermometers for HVAC

6

6. Indicator check

The check of the indicator should only be carried out in comparison with a more

accurate instrument or, if possible, with a calibrated instrument. The temperature

during the check must remain constant. Fluctuating temperatures may lead to

reading errors caused by the different response times of the probes.

For thermometers without protective sleeves, a minimum waiting time of

10 minutes, for temperature equalisation, should be observed, provided the

appropriate immersion length is adhered to.

Attention: Only gaseous pressure-free media are admissible!

For thermometers with protective sleeves, a minimum waiting time of at least

15 minutes, for temperature equalisation, should be observed, provided the

appropriate immersion length is adhered to.

Admissible working pressure at the pocket: max. 6 bar

7. Indicator correction

Any interference with or modification to the instrument will invalidate

the warranty!

The indicator correction built into the bottom of stem should only be operated

using a screwdriver, and only if the thermometer has been damaged due to

improper handling, severe shocks, during transport, etc.

When correcting an indication error using the adjustment mechanism, a

calibrated thermometer must be used for comparison. Subsequently, the

thermometer must be checked at several temperature points.