beautypg.com

WIKA 233.50 User Manual

Page 3

background image

Permissible vibration load at the installation site

„

The instruments should always be installed in locations free from vibration.

„

If necessary, it is possible to isolate the instrument from the mounting point by instal-

ling a flexible connection line between the measuring point and the pressure gauge and

mounting the instrument on a suitable bracket.

„

If this is not possible, the following limit values must not be exceeded:

Dry gauges:

Frequency range < 150 Hz

(Model 232)

Acceleration < 0.7 g (7 m/s

2

)

Liquid-filled gauges: Frequency range < 150 Hz

(Model 233)

Acceleration < 4 g (40 m/s

2

)

The liquid filling must be checked on a regular basis.

The liquid level must not drop below 75 % of the gauge diameter.

4. Commissioning

During the commissioning process pressure surges must be avoided at all costs. Open the

shut-off valves slowly.

5. Maintenance and cleaning

„

The instruments are maintenance-free.

„

The indicator and switching function should be checked once or twice every year. The

gauge must be disconnected from the process to check with a pressure testing device.

„

Clean the pressure gauge with a moist cloth.

„

Repairs must only be carried out by the manufacturer or appropriately qualified skilled

personnel.

„

When dismounting, close the compensating valve (if available).

WARNING!

Residual media in dismounted pressure gauges can result in a risk to persons,

the environment and equipment.

Take sufficient precautionary measures.

Appendix 1: Declaration of conformity

This manual is related to the following products: