WIKA operating instructions air2guide models A2G-10, A2G-15

GB

8

40197310.04 04/2012 GB/D/F/E/I

5. Transport, packaging ... / 6. Commissioning, operation

5. Transport, packaging and storage

5.1 Transport

Check the pressure gauge for any damage that may have been caused

by transport. Obvious damage must be reported immediately.

5.2 Packaging

Do not remove packaging until just before mounting.

Keep the packaging as it will provide optimum protection during trans-

port (e.g. change in installation site, sending for repair).

5.3 Storage
Permissible conditions at the place of storage

Storage temperature: -30 ... +80 °C

Protect the instruments from moisture and dust.

6. Commissioning, operation

Installation and mechanical connection
In accordance with the general technical regulations for pressure

gauges (e.g. EN 837-2 "Selection and installation recommendations for

pressure gauges").

Process connection lower mount (LM) or back mount (BM)

Protect measuring instruments from contamination, high temperature

changes and vibrations

air2guide standard gauges are calibrated in vertical position and

should be installed in the same position to avoid loss of class

accuracy. If a different installation position is required please specify

when ordering. Gauges with high pressure ranges can be installed in

different non-vertical mounting position, by simply adjusting the zero

point.