WIKA 532.53+8xx User Manual
Page 7

7
WIKA Operating instructions pressure gauges Model 5 with 831 per ATEX
2133086 08/2004 GB/D/F
GB
Permissible vibratory stress at the mounting location
T
As a matter of principle the instruments should only be mounted at locations
without vibratory stresses
T
Where required, a decoupling from the mounting location can be achieved
e.g. by a flexible connecting line from the measuring point to the pressure
gauge and mounting via a measuring instrument bracket.
T
If this is not possible, the following limits must not be exceeded:
Dry gauges:
Frequency range < 150 Hz
(Models 532.XX)
Acceleration < 0.7 g (7 m/s
2
)
Liquid-filled gauges:
Frequency range < 150 Hz
(Model 533.XX)
Acceleration < 4 g (40 m/s
2
)
The liquid filling has to be checked on a regular basis.
The liquid level must not drop below 75% of the gauge diameter.
4. Alarm contacts
EC-type examination certificates
T
Standard version Model 831.XX
PTB 99 ATEX 2219 X (enclosure 2) and ZELM 03 ATEX 0128 X (enclosure 4)
Depending on the number of switches and on the case diameter either
Model SJ2-N... or Model SJ3.5-...-N... is used.
T
Safety pattern version Models 831.XX - SN or - S1N
PTB 00 ATEX 2049 X (enclosure 3) and ZELM 03 ATEX 0128 X (enclosure 4)
Depending on the number of switches and on the case diameter either
Models SJ 2-SN..., SJ 2-S1N..., SJ 3.5-SN.. or SJ 3.5-S1N... are used.
The built-in sensor type is stated on the product label of the pressure gauge.
Wiring details
T
The electrical connections should be made by qualified electricians
T
Connection of the switches via screw terminals in the terminal box
T
Conductor cross section max. 1.5 mm
2
T
The terminal assignment is stated on the connection plate at the
pressure gauge
3. Technical data ... / 4. Alarm contacts