Troubleshooting – BrandTech Scientific macro User Manual

Page 31

background image

English

31

Troubleshooting

Trouble

Possible Cause

Action to be taken

Suction capacity
impaired

Filter dirty or valve
system damaged

Replace filter,
see pages 26-27

Pipette drips

Filter not properly posi-
tioned or valve system
damaged

Insert filter properly or
replace valve system,
see page 28

Pipette not held
properly

Silicone adapter
damaged

Unscrew the adapter
support of the instru-
ment, replace silicone
adapter

Compressed suction
bellows take in air
automatically

Suction bellows or valve
system damaged

Replace the looking ring
with the suction bellows
or the valve system

Often visual control of the instrument is sufficient to discover a faulty
device which can easily be replaced by a spare part. The advantages of
do-it-yourself repair are obvious: It saves time and money.