Cleaning · autoclaving – BrandTech Scientific macro User Manual

Page 27

background image

8

6

English

27

8.

The entire instrument incl. membrane

filter is autoclavable at 121 °C (2 bar)
according to DIN EN 285.

For autoclaving remove suction bellow.

6.

Place the new membrane filter with

the thicker end facing downwards into the
adapter (with the smaller conical side up).

Not every membrane filter is suitable. Only use genuine manufacturer’s
recommended accessories. Filters of a pore size of 3 µm are a standard
equipment of the instrument. The membrane filter is autoclavable up
to 5 times at 121 °C. It is the user's responsibility to ensure effective
autoclaving.

Note:

Cleaning · Autoclaving

7.

Screw in the adapter support hand tight.