Troubleshooting, English – BrandTech Scientific Dispensette TA User Manual
Page 45

English
45
Liquid is above the
piston
Piston is leaking.
Perform a standard cleaning, replace the piston/
cylinder assembly (see page 39-41).
Piston difficult
to move
Formation of crystals, dirty
Stop dispensing immediately. Follow all cleaning
instructions (page 39-41).
Filling
not possible
Volume adjusted to
minimum setting
Set to required volume (see page 33).
Filling valve stuck
Clean the filling valve.If the valve is stuck use a
200 µl pipette tip to loosen it (see page 40). If
necessary, replace the filling valve and sealing
washer.
Dispensing
not possible
Discharge valve stuck
Clean discharge valve. If necessary replace dis-
charge tube (see page 40).
Air bubbles in
the instrument
Reagent with high vapor
pressure has been drawn in too
quickly
Draw in reagent slowly.
Seal not inserted, knurled
locking nuts not firmly
connected
Confirm that the seal is in place, and that the
knurled locking nuts at the discharge tube and the
SafetyPrime
™
recirculation valve are firmly seated
and finger tight.
The instrument has not
been primed
Prime the instrument (see page 32).
Filling tube is loose or
damaged
Push the filling tube on firmly. If necessary, cut
off approx. 1 cm of tube at the upper end and re-
connect it or replace filling tube.
Valves not firmly connected
or damaged
Follow cleaning procedure (see page 39, 42).
Tighten the valves using the mounting tool. If nec-
essary, replace the valves and sealing washers.
Recirculation tube not
connected
Connect recirculation tube (see page 31, Fig. 5).
Dispensed volume
is too low
Discharge tube is loose
Push the discharge tube on firmly and finger
thighten locking nuts.
Filling tube is loose
or damaged
Cleaning procedure (see page 39-40). Push the
filling tube on firmly. If necessary, cut off approx.
1 cm of the tube at the upper end and re-connect
it or replace filling tube (see page 40).
Filling valve is loose, con-
taminated, or damaged
Cleaning procedure (see page 39-40). Tighten
the valves using the mounting tool. If necessary,
replace filling valves and sealing washers.
Liquid leaking at the
safety ring
Dispensing cartridge is loose,
or the piston seal is damaged
Tighten the safety ring, and if necessary replace
the dispensing cartridge
Leaking liquid
between instrument
and bottle
Recirculation tube not con-
nected
Connect recirculation tube (see page 31, Fig. 5).
Volatile reagent dispensed
without sealing ring
Mount sealing ring.
Problem Possible cause Corrective action
Troubleshooting