beautypg.com

Myron L 4P (includes Addendum 10-01) User Manual

Page 7

background image

B. Measuring Resistivity

Resistivity is for low conductivity solutions. In a cell cup the value may drift
from trace contaminants or absorption from atmospheric gasses, so
measuring a flowing sample is recommended.

1.

Hold instrument at 30° angle (cup sloping downward).

2.

Let sample flow continuously into conductivity cell with no
aeration.

3.

Press key; use best reading.

NOTE: If reading is lower than 10 kilohms, display will be dashes: [ - - - - ].
Use Conductivity.

V.

SOLUTION SELECTION

A. Why Solution Selection is Available

Conductivity, Resistivity, and TDS require temperature correction to 25°C
values (ref. Standardized to 25°C, pg. 23). Selection determines the
temperature correction of conductivity and calculation of TDS from
compensated conductivity (ref. Cond. Conversion to TDS, pg. 26).

B. The 4 Solution Types

On the left side of the display is the salt solution characteristic used to
model temperature compensation of conductivity and its TDS
conversion. Generally, using KCl for conductivity, NaCl for resistivity and
442 (Natural Water characteristic) for TDS will reflect present industry
practice for standardization. This is the setup as shipped from the factory
(ref. Solution Characteristics, pg. 26).
The USER selection allows a custom value to be entered for the
temperature compensation of conductivity and also the conversion ratio if
measuring TDS.

C. Calibration of Each Solution Type

There is a separate calibration for each of the 4 solution types. Note that
calibration of a 442 solution does not affect the calibration of a NaCl
solution. For example: Calibration (ref. Conductivity or TDS Calibration,
pg. 14) is performed separately for each type of solution one wishes to
measure (ref. Conductivity/TDS Standard Solutions, pg. 22).

D. Procedure to Select a Solution

N O T E

: Check display to see if solution displayed (KCl, NaCl, 442 or

USER) is already the type desired. If not:
10

R E S

1.

Press ,

or to select the parameter on which

you wish to change the
solution type.

2.

Press and hold key

about 3 seconds to make

S E L

” appear (see Figure 1).

(For demonstration purposes,
all 4 solution types are shown simultaneously.)

3.

Use

or

key to obtain type of solution desired (ref.

Solution Characteristics, pg. 26). The selected solution type will
be displayed: KCl, NaCl, 442 or User.

4.

Press to accept new solution type.

E. Application of USER Solution Type

1. User Programmable Tempco

This feature allows you to change your Ultrameter’s temperature

compensating factor to another factor between 0-9.99%/°C (ref.

Temperature Compensation, pg. 23).

a.

As in Procedure to Select a Solution, pg. 10, select “USER”
mode.

b.

With “USER” mode now selected, press . You may now

adjust a temperature compensation from .00%/°C to 9.99%/°C,

by pressing

or .

See example in Figure 2.

c.

Press twice to skip

calibration adjustment and
accept the new tempco (3
times if in TDS mode). You are now ready to measure samples
with your new temperature compensation factor.

11

M S

MR

CAL

MCLR

CAL

MCLR

R E S

COND

T D S

CAL

MCLR

M S

MR

CAL

MCLR

Figure 2

COND

°C

% /

User

Figure 1

KCl

NaCl

User

4 4 2