SPECIFICATIONS
MYRON L
®
COMPANY
2450 Impala Drive • Carlsbad, Ca 92010-7226 • Phone: +1-760-438-2021
E-Mail: Customer Service – [email protected] • Technical Support – [email protected]
Website: www.myronl.com
MADE IN USA
©2014 MYRON L
®
COMPANY
Do not over tighten.
App Requirements:
Apple iOS 6 or iOS 7. Optimized for iOS 7.
Measurement Range:
0.00 – 14.00 pH; Temperature: 0°-71° C/32-160° F
Accuracy (After Wet Calibration):
± 0.01 pH; Temperature ± 0.1ºC / ± 0.1ºF
Repeatability:
± 0.01 pH; Temperature ± 0.1ºC / ± 0.1ºF
Resolution:
0.01 pH; Temperature ± 0.1ºC / ± 0.1ºF
Time to Reading Stabilization:
10 - 30 seconds
Active Mode Power Consumption: Active Mode 137 mA, Sleep Mode 2 µA
Temperature Compensation:
Automatic to 25ºC
Physical Dimensions:
17.15 cm L x 1.59 cm D or 6.75 in. L x .625 in. D
Weight:
54 g or 1.94 oz
Case Material:
Anodized Aircraft Aluminum with Protective Coating
Battery Type:
N type, Alkaline 1.5 V
Calibration Standard Solution:
4.0 pH, 7.0 pH, 10.0 pH
Operating/Storage Temperature:
0 - 55ºC / 32 - 131ºF
Water Resistance:
IP67 and NEMA 6
EN61236-1: 2006 - Annex A:
2008; 1999/5/EC Annex III
Electrostatic discharge to case of instrument may
cause PTBT2 to spontaneously power on. In this case,
the PTBT2 will power off after several seconds
FCC ID: T7VPAN17
The Bluetooth transceiver device meets the requirements for
modular transmitter approval as detailed in FCC public Notice
DA00-1407.
Canada (IC), license:
IC: 216Q-PAN17
The Bluetooth transceiver device meets the requirements for
modular transmitter approval as detailed in RSS-GEN.
ORIGINAL
EDITED FOR USE
Minimum height 5 mm
QUICK REFERENCE INSTRUCTIONS
1. Turn on ULTRAPEN App: Tap ULTRAPEN icon on mobile device’s home screen.
2. Turn on PTBT2: Press and release push button on pen.
3. LED flashes rapidly: Dip pen in sample solution so cell is totally submerged.
4. LED flashes slowly: Swirl pen to remove bubbles, keep cell submerged.
5. LED turns solid ON then turns OFF: Measurement is complete and reading is displayed.
Feature Navigation Bar
This button always returns the App to the Measurement Screen.
This button displays a list of saved measurements.
This button places the App in Calibration mode.
This button opens the App’s Bluetooth Connect / Disconnect screen.
This button opens the App’s preferences and settings mode.
An internet link to full Operations Manual (Help button) is found here.
ACCESSORIES
STANDARD SOLUTIONS FOR CALIBRATION: The ULTRAPEN™ PT2 requires pH 4, pH 7, and pH 10
buffer solutions for wet calibration and pH Sensor Storage Solution for proper storage.
— Order MODEL#s: PH4, PH7, PH10, and PHSS.
Replacement pH Sensor (with instructions) — Order Model: RPT2
.
III. pH Sensor Replacement
(Follow the instructions that come with your Replacement Sensor.)
Only Remove/Replace the pH sensor in a CLEAN and DRY environment!
To remove the pH sensor:
1. Remove the soaker cap; make sure the PTBT2 (including the pH sensor) is clean
and dry. Loosen the battery tray to allow pressure equalization.
2. Then firmly grasp the pH sensor body and slowly pull the pH sensor out.
To install a new pH sensor:
1. Line up the alignment tabs on the pH sensor with the alignment slots on the PT2
unit.
2. Gently push the pH sensor into position, close battery cap.