beautypg.com

A. calibration preparation, B. ph calibration, C. orp calibration – Myron L PT4 User Manual

Page 4: V. solution check

background image

1

However, you should check the calibration whenever measurements are not as expected.

For greatest accuracy, you should perform a 3-point wet pH calibration, and wet ORP

calibration with the ORP Standard Solution closest in value to the solution you will be

testing.

NOTE: If the measurement is NOT within calibration limits for any reason, “Error” will

display. Check to make sure you are using a proper Myron L Company pH Buffer or ORP

Standard Solution. If the solution is correct, clean the sensor as described in Sensor

Cleaning section on page 4. Restart calibration.

NOTE: Small bubbles trapped in the sensor may give a false calibration. After calibration

is completed, measure the pH Buffer or ORP Standard Solutions again in solution check

mode “SOL ck” (see pages 3 and 4) to verify correct calibration.

NOTE: If at any point during calibration, you do not submerge the sensor in solution

before the flashing slows, allow the

PT4

to power off and start over.

NOTE: You should always calibrate with pH 7 first.

A. Calibration preparation

For maximum accuracy, fill 2 clean containers with each pH Buffer and/or ORP Standard

Solution. Arrange them in such a way that you can clearly remember which is the rinse

solution and which is the calibration standard/buffer. If you don’t have enough standard/

buffer, you can use 1 container of each standard/buffer for calibration and 1 container of

clean water for all rinsing. Always rinse the FC

E

sensor between standard/buffer solutions.

Ensure the FC

E

sensor is clean and free of debris.

B. pH Calibration

using Myron L Company pH 7, 4, and 10 Buffer

Solutions.

NOTE: You should always calibrate with pH 7 first.

1. Thoroughly rinse the

PT4

by submerging the sensor in pH 7 Buffer rinse solution and

swirling it around.

2. Push and release the push button to turn the

PT4

on.

3. Push and hold the push button. The display will alternate between “CAL”, “FAC CAL”,

“ºCºF TEMP”, “ModE SEL”, “PAr SEL”, “SOL ck”, and “ESC”.

4. Release the button when “CAL” displays.

5. The display will alternate between “PUSHnHLD” and “CAL.

6. Push and hold the button, The display will alternate between “PH” and “ORP”.

7. Release the button when “PH” is displayed.

8. The display will indicate “CAL” and the LED will flash rapidly.

9. While the LED flashes rapidly, dip the

PT4

in pH 7 Buffer Calibration Solution so that

the sensor is completely submerged.

10. While the LED flashes slowly, the pH calibration point will display along with “CAL”.

Swirl the

PT4

around to remove bubbles, keeping the sensor submerged.

11. If the pH 7 calibration is successful, the display will indicate “SAVEd”, then

“PUSHCONT” will be displayed (“PUSHCONT” will NOT be displayed if only calibrated

with pH 4 or 10).

12. Push and release to continue or let the unit time out to exit after a 1-point or 2-point

calibration.

13. Repeat steps 9 through 12 with pH 4 and 10 Buffer Solutions. After the 3rd calibration

point is successfully saved, the display will indicate “SAVEd” and power off.

14. Verify calibration by retesting the calibration solution in solution check mode “SOL

ck”, see section V below.

C. ORp Calibration

using Myron L Company 80mV Quinhydrone, 260mV

Quinhydrone, or 470mV MLC Light’s ORp Standard Solution.

N

OTE: The PT4 has automatic temperature compensation in ORP calibration mode

(from 15ºC to 30ºC).

1. Follow pH calibration steps 1 through 6, using ORP Solutions.

2. Release the button when “ORP” is displayed.

3. The display will indicate “CAL” and the LED will flash rapidly.

4. While the LED flashes rapidly, dip the PT4 in ORP Standard Solution so that the

sensor is completely submerged.

5. While the LED flashes slowly, the ORP calibration point will display along with “CAL”.

Swirl the

PT4

around to remove any air bubbles, keeping the sensor submerged.

6. If the ORP calibration is successful, the display will indicate “CAL SAVEd”, then

time out.

7. Verify calibration by retesting the calibration solution in solution check mode.

V. SOLUTION CHECK

Solution check is provided to verify the proper calibration value was recorded when

using Myron L Company pH Buffers and ORP Standard Solutions. To verify proper

calibration, simply put the

PT4

into solution check mode, select the mode to verify (pH

or ORP), then dip the sensor into the pH Buffer or ORP Calibration Solution so that the

sensor is completely submerged and swirl around to release any air bubbles, then verify

displayed value matches the value on the bottle.

3

Solution Check continued on page 4 ...