beautypg.com

Iv. calibration, A. calibration preparation, B. calibration – Myron L PT3 User Manual

Page 4: C. factory calibration

background image

1

a. In Hold mode when the LED turns on solid, remove the PT3 from

solution. The display will alternate between the final ORP and temperature

readings. Note the readings for your records.

b. In LIVE mode allow the PT3 to remain in solution while the LED flashes

slowly. The display will alternate between live ORP and temperature

readings. Note the readings for your records. LIVE measurement will time

out after 90 seconds OR push and release the push button to turn the

PT3 off at any time during LIVE measurement.

IV. Calibration

The Myron L Company recommends calibrating twice a month, depending on

usage. However, you should check the calibration whenever measurements are

not as expected.

The PT3 has automatic temperature compensation in calibration mode (from 15ºC to

30ºC). Temperature affects the reaction potentials for all chemicals differently. True

ORP is the direct measurement of electron activity during an oxidation-reduction

reaction, regardless of temperature. However, for maximum accuracy and ease of

calibration, the Myron L Company has developed three calibration solutions with

known dissolved species and embedded automatic temperature compensation for

those solutions into the calibration function of your PT3. To verify calibration while in

measurement mode, you must manually correct for temperature variations. For more

information, visit www.myronl.com/pt3.htm.

NOTE: If the measurement is NOT within calibration limits for any reason, “Error”

will display. Check to make sure you are using a proper ORP Standard Solution.

If the solution is correct, clean the sensor as described on page 4 under Routine

Maintenance. (Reconditioning the sensor should not be necessary due to the high

ionic strength of the calibration solutions.) Restart calibration.

NOTE: Small bubbles trapped in the sensor may give a false calibration. After

calibration is completed, measure the ORP Standard Solution again to verify

correct calibration. Remember, in measurement mode you must manually correct

for temperature variance from 25ºC. Example: At 25ºC, ORP2602OZ Standard

Solution will read 260mV; however, at 20.0ºC ORP2602OZ will read 265mV.

NOTE: If at any point during calibration, you do not submerge the sensor in solution

before the flashing slows, allow the PT3 to turn off and start over.

A. Calibration Preparation

1. Ensure the ORP sensor is clean and free of debris.

2. For maximum accuracy, pour a small amount of the calibration solution into

a separate container to rinse the sensor in prior to calibration. If this is not

possible, rinse the sensor in clean water prior to calibration.

B. Calibration

Use Myron L Company 80mV Quinhydrone, 260mV Quinhydrone, or 470mV

MLC Light’s ORP Standard Solution.

1. Thoroughly rinse the PT3 by submerging the sensor in ORP calibration rinse

solution and swirling it around.

2. Push and release the push button to turn the PT3 on.

3. Push and hold the push button. The display will alternate between “CAL”, “FAC

CAL”, “ºCºF TEMP”, “ModE SEL” and “ESC”.

4. Release the button when “CAL” displays. The display will indicate “CAL” and

the LED will flash rapidly.

5. While the LED flashes rapidly, dip the PT3 in ORP Standard Solution so that

the sensor is completely submerged.

6. While the LED flashes slowly, the display will indicate “ [value] CAL”; swirl the

PT3 around to remove bubbles, keeping the sensor submerged.

7. If the ORP calibration is successful, the display will indicate “CAL SAVEd”, then

time out.

8. Verify calibration by retesting the calibration solution.

C. factory Calibration

When ORP Standard Solution is not available, the PT3 can be returned to factory

default calibration using the FAC CAL function. This will erase any stored wet

calibration.

NOTE: default factory calibration resets the electronics only and does NOT take

the condition of the sensor into consideration.

To return your PT3 to factory calibration:

3

continued on page 4 ...