Accessories continued, Warranty/service, Range extender – Myron L 532M1 User Manual
Page 4: Porta pak, Ordering

The Myron L 532T1, 532T2, 532M1 Meters have a limited two
year warranty. If your instrument fails to operate properly, check
the battery and calibration. If it still fails to function properly,
return it prepaid to the Myron L Company.
Faulty instruments may be returned to us without prior
authorization.
meters within two year warranty period:
Failures due to materials or workmanship will be repaired or
replaced (our option) without charge if returned freight prepaid.
If failure is deemed by the factory to have been caused by abuse
or tampering, the following procedure will apply.
instruments/controls out of warranty:
Diagnosis will be made and repairs completed, providing the
repair charges are $80.00 or less.
note: Actual repair charges may be less than this amount.
we will diagnose (but not repair) a returned meter and supply
an estimate of charges if AnY of the following apply:
1. Repair charges will be more than $80.00.
2. You specifically request an estimate of required repairs and
charges.
3. The cost of required repairs exceeds one-half the list price of
a new instrument.
4. The instrument is over ten years old. Because of component
changes and improvements, such instruments can no longer
be repaired.
note: unrepaired meters are discarded unless you want them
returned to you. If so, there is currently a $35.00 charge per
unrepaired instrument to cover diagnosis and handling.
This warranty is limited to the repair or replacement of the
Myron L 532T1, 532T2, 532M1 Meters only. The Myron L
Company assumes no other responsibility or liability.
warranty/service
2450 Impala Drive
Carlsbad, CA 92010-7226
TELEpHonE: +1-760-438-2021
FAx: +1-760-931-9189
www.myronl.com
© Myron L Company 2008
3RoM wEB 05-08
Specifications subject to change without notice
printed in the u.S.A.
recommended standard solutions
model/range
solution number
532t1
x 10
442-30
x 100
442-300
x 1000
442-3000
with RE-10
442-30,000
532t2
x 10
442-15
x 100
442-150
x 1000
442-1500
with RE-10
442-15,000
532m1
x 10
442-30
x 100
442-300
x 1000
442-3000
with RE-10
442-30,000
range extender
The RE-10 Range Extender is a useful accessory for
testing high conductivity/ppm solutions beyond the normal
range of your DS Meter™. Inserting the RE-10 into a
sample-filled cell cup increases the maximum range ten
times. use the Range Extender whenever the reading is
off the scale.
using the re-10
1. Fill the cell cup three times to rinse it, but each time
insert the Range Extender to rinse it also.
2. Fill the cell cup with your sample. push the Range
Extender into the cell cup, seating the o-ring seal.
3. use and read the DS Meter in the normal manner.
Multiply the reading by 10.
4. For best accuracy, repeat the complete test with a
fresh sample.
5. when you’re done testing, remove the Range
Extender. Thoroughly rinse the cell cup and Extender
with clean water (preferably distilled or deionized) to
eliminate dried salts build-up. This is extremely
important when the instrument will be used to test high
purity water.
calibrating the re-10
1. Calibrate x1000 range of meter (without RE-10) with
Standard Solution
2. Fill cell cup with appropriate high conductivity
Standard Solution (442-15,000 or 442-30,000).
3. Insert RE-10 and press the button. Multiply the reading
by 10 and compare it with the value on the Standard
Solution label. It they are not the same the RE-10 must
be recalibrated.
note: Do noT adjust the Conductivity Calibration
Control.
4. Adjust the white insert of the Extender as follows:
If the reading is too high — push or tap inward.
If the reading is too low — twist or pull outward with pliers.
note: The Range Extender is factory calibrated to a
particular meter. It should be recalibrated if it is to be used
with another meter.
porta pak
Carrying Case for use with
all Myron L DS/pDS meters.
Foam lined and molded of
sturdy ABS plastic
ordering
To order accessories contact your nearest stocking
distributor, or the Myron L Company.
RE-10 Range Extender
accessories continued