Maintenance, Preventive maintenance calendar, Maintenance preventive maintenance calendar – Arizona Hydrogen Sulfide Analyzer Jerome 631-X User Manual
Page 18

5. MAINTENANCE
Preventive Maintenance Calendar
To keep the Jerome 631-X operating at peak performance, follow the maintenance schedule below
as a guide. Since maintenance is more a function of application and amount of use rather than
time, your requirements may be different from the listed schedule. Call AZI Customer Service at
800-528-7411, 602-470-1414, or e-mail [email protected] for additional guidance for your
environment and operation.
PART/COMPONENT
MAINTENANCE CYCLE
REFER TO
PAGE
Charge batteries
At least once per month, after 1 month's storage,
or when LO BAT appears
Change .25 inch fritware
Weekly or as needed
Change internal filters and
tubing
After 6 months of use or as needed
Replace zero air filter
Factory calibration
Annually
Calibration check
Monthly or as needed
Appendix
A, Page 36
Replace batteries
Annually or as needed.
The battery pack contains NiCad batteries.
Dispose of properly.
NOTE: Install the zero air filter into the instrument's intake during storage.
1
Zero air filters, LFS and LFD scrubber filters contain Resisorb™. Call AZI Customer Service at 800-528-7411,
601-470-1414, or e-mail [email protected] for a copy of the Material Safety Data Sheet. Dispose of all filters
properly.
Page 18 of 50
AZI Customer Service 800-528-7411 or 602-470-1414