Multi-station patient room applications – Aiphone 83873900 0602 E User Manual
Page 22
![background image](/manuals/42765/22/background.png)
– 22 –
WIRING DIAGRAMS
Multi-Station Patient Room Applications
1. Four bed application covered by four NH-1SA/A sub stations
with a common bathroom pullcord
Multiple NH-1SA/A and NH-2SA/A bedside sub stations can also be installed
within a patient room. Observe the following conditions;
• Max. 6 bedside sub stations may be connected per room for max. 12 calls (when NH-2SA/A stations are used).
• A corridor lamp inside (NHR-4A) must not be connected with more than 4 bedside sub stations.
• Due to the above, install two NHR-4A corridor lamps when more than 4 sub stations are in one
room, i.e. two WHITE lamps used in one NHR-4A, and one RED bulb used in the other NHR-4A.
• NH-1SA/A and NH-2SA/A station numbers must be programmed in numerical order and
connected to the same subs trunk line.
• NHR-7A bathroom pullcord is optional.
NH-1SA/A
NH-1SA/A
NH-1SA/A
NH-1SA/A
1
2
3
4
5
6
WL
RL
E
TC
TL
E
ZL
LC
1
2
3
4
5
6
WL
RL
E
TC
TL
E
ZL
LC
1
2
3
4
5
6
WL
RL
E
TC
TL
E
ZL
LC
1
2
3
4
5
6
WL
RL
E
TC
TL
E
ZL
LC
Z1
11
12
13
14
15
16
3P(or 4P)
3P
2
3P(or 4P)
3 or 4
3
3P
2
3P
2
NHR-4A
NHX-80X
CCU
NHR-7A
(common
bathroom)
NHR-4A
NHR-7A
Brown
Orange
Red
Orange
Brown
Red
Orange
Yellow
Green
Blue
Purple
Brown
Orange
Purple
Gray
Brown
Orange
Purple
Gray
Brown
Orange
Purple
Gray
Yellow
Green
Blue
Blue
(*1)
(*2)
CN1
CN2
To Subs
block 1
Each pair: polarized
TO next patient room
Subs trunk line 1
The above example shows a patient room with four beds, in which a common bathroom pullcord is used.
Multiple sub applications can be arranged in the following manner.
Beds in Room
2
3
4
5
6
Subs combinations
One NH-2SA/A or two NH-1SA/A
One each NH-2SA/A & NH-1SA/A
or three NH-1SA/A
Two NH-2SA/A or four NH-1SA/A
Two NH-2SA/A & one NH-1SA/A
Three NH-2SA/A
Beds in Room
7
8
9
10
11
12
Subs combinations
Three NH-2SA/A & one NH-1SA/A
Four NH-2SA/A
Four NH-2SA/A & one NH-1SA/A
Five NH-2SA/A
Five NH-2SA/A & one NH-1SA/A
Six NH-2SA/A
(*1) Terminal Z1 for lighting Lamp 1 in NHR-3A-4.
(*2) NHR-7A may be wired to any sub station convenient. Terminal RL is wired
only on the sub station which NHR-7A is connected to.
Note; Bedside sub station Models: NH-1SA and NH-2SA CANNOT be used in multiple bed patient rooms.