Wiring, Datalogger programming – Campbell Scientific SP-LITE Solar Radiation Sensor User Manual
Page 11

SP-LITE Silicon Pyranometer
7
4. Wiring
The SP-LITE can be measured with a differential or single-ended channel on
the datalogger; a differential channel is recommended.
FIGURE 8. SP-LITE to Datalogger Connections
TABLE 1. Datalogger Connections for Differential Measurement
Function
Color
CR10X,CR510
CR23X
21X/CR7
CR1000
Radiation Signal
White
Differential H
Differential H
Differential H
Differential H
Signal Reference
Green
Differential L
Differential L
Differential L
Differential L
Jumper to
AG
Shield
Clear
G
TABLE 2. Datalogger Connections for Single-Ended Measurement
Function
Color
CR10X,CR510
CR23X
21X/CR7
CR1000
Radiation Signal
White
S.E. Channel
S.E. Channel
S.E. Channel
S.E. Channel
Signal Reference
Green
AG
Shield
Clear
G
5. Datalogger Programming
The SP-LITE outputs a low-level voltage ranging from 0 to a maximum of
about 15mV. A differential voltage measurement (Instruction 2, VoltDiff) is
recommended because it has better noise rejection than a single-ended
measurement. If a differential channel is not available, a single-ended
measurement (Instruction 1, VoltSE) can be used. The acceptability of a single-
ended measurement can be determined by simply comparing the results of
single-ended and differential measurements made under the same conditions.
Radiation Signal
White
Signal Reference
Green
Shield
Clear
(User supplied jumper wire for differential measurement only)