SC105 CS I/O to RS-232 Interface

RS-232 Baud Rates:

The SC105 will support the following baud
rates: 1200, 9600, 19200, 38400, 57600, and
115200

RS-232 Parity and Data
The SC105 supports the following:

Parity:

Even, Odd, None

Data Bits: 7,

8

CS I/O Port Modes:

CSDC, SDC, ME, Addressed Print Device for
P96 output

Electrical:

The SC105 uses power from the +5 V line on
the 9-pin interface connected to the datalogger

Current

Standby: 0.16

mA

Communicating:

1 to 4 mA

Additional current (up to 8 mA) from the 5 V
supply may be used by the RS-232 device
connected to the SC105

Physical

Height:

2.3 cm (0.9 in)

Width:

4.1 cm (1.6 in)

Length:

8.9 cm (3.5 in)

Weight:

45.4 g (1.6 oz)

Environmental

Temperature:

–25° to +50°C

Humidity:

up to 95% non-condensing

3. Set-up Menu

The SC105 has a built-in set-up menu for configuring communication mode,
CS I/O port configuration, RS-232 port configuration, and other parameters.
The Set-up Menu is shown in FIGURE 3-1. It is accessed by connecting the
SC105’s RS-232 port to a PC with the included null modem cable. The SC105
also needs to have power. Usually, the SC105 is powered by connecting the
SC105’s CS I/O port to the CS I/O port of a datalogger.

With the null modem connection to the PC, typically Campbell Scientific’s
Device Configuration Utility (DevConfig) is used, but a terminal program such
as HyperTerminal

TM

or Procomm

TM

(always 9600 baud, 8-N-1) can also be

used. Press the Program button on the SC105 for one second to access the
Set-up Menu. Changed settings are saved in flash memory by selecting Apply
in DevConfig or menu item 4 if using HyperTerminal or Procomm. If left idle,
the Set-up Menu will time out 60 seconds after the last received character and
exit without saving any parameter changes with the message “Set-up Timeout.”
A datalogger can remain connected to the CS I/O port (to provide power to
SC105) while setting SC105 parameters on the RS-232 port, although CS I/O
communications would be inactive until exiting the Set-up Menu.

3