CS115 Barometric Pressure Sensor

8

8: Sample (P70)

1:

1

Reps

2:

1*

Loc [ P_mb ]

*Table 2 Program

02: 0.0000

Execution Interval (seconds)

*Table 3 Subroutines

End Program

-Input Locations-
1 P_mb

* Proper entries will vary with program and datalogger channel, and input location assignments.

5.1.2 Period Averaging Measurement Example for CR10(X) in a Slow

Executing Program

EXAMPLE 2. Sample Program for CR10(X)

;{CR10X}
;
*Table 1 Program

01: 10

Execution Interval (seconds)

;Turn on CS115
;
1: Do (P86)

1:

48*

Set Port 8* High

;Give 2 second delay for CS115 to settle
;
2: Excitation with Delay (P22)

1:

1

Ex Channel

2:

0

Delay W/Ex (0.01 sec units)

3:

200

Delay After Ex (0.01 sec units)

4:

0

mV Excitation

3: Period Average (SE) (P27)

1:

1

Reps

2:

14

200 kHz Max Freq @ 2 V Peak to Peak, Freq Output

3:

6*

SE Channel

4:

10

No. of Cycles

5:

5

Timeout (units = 0.01 seconds)

6:

1*

Loc [ P_mb ]

7:

1000

Mult

8:

0.0

Offset