Aalborg Vortex User Manual
Page 87

82
COMMAND
NAME
DESCRIPTION
NO.
COMMAND SYNT
A
X
COMMAND
ARGUMENT 1
ARGUMENT 2
ARGUMENT 3
ARGUMENT 4
RESPONSE
Meter
Info
Requests meter
configuration info:
-
full scale range
(L/min)
-
Meter size (inches)
-
T/P option support
(Y
,N)
-
4-20 mA Output #1
configuration
-
4
-20 mA Output #2
configuration
7
MI
MI: 1000.0,0.75,Y
,V
,M
1000.0 – full scale L/min
0.75 – meter size [inch]
Y –
T/P supported
N – T/P not supported
V –
volumetric flow
M –
mass flow
T –
Temperature
P –
P
ressure
D –
Density
O –
Off (Disabled)
Flow
Alarms
Sets / reads the status
of the flow alarms.
Note: High and Low limits
for V
o
lumetric flow
configuration have to be
entered in fraction %FS
notation (0.0 – 1.0). High
and Low limits for Mass
flow configuration have to
be entered in the currently
active mass flow units.
High alarm value has to
be more than Low alarm
value.
Alarm conditions:
Flow > High Limit = H
Flow < Low Limit = L
Low < Flow < High = N
8
FA
C (flow configuration)
V
-volum. flow
M-mass flow
alue> (high limit) alue> (low limit) FAC:V ,0.9,0.1 A (action delay in sec) alue> [0-3600] FAA: alue (sec)> E (enable alarm) FA :E D (disable alarm)* FA :D R (read current status) FA R:N (no alarm) FA R:H (high alarm) FA R:L (low alarm) S (Read current settings) FAS:M,C,H,L,D,B where: M – mode (E/D) C– Configuration (V/M) H – High settings value L – Low settings value D – Action Delay (sec) B – Latch mode (0-1) L ( L at ch m o d e) alue> (0-disabled*) (1-enabl’d) FA L: alue> where: V alue = 0 – 1