Smode, Intv – Vaisala HMP155 User Manual
Page 34

USER'S GUIDE____________________________________________________________________
32 _______________________________________________________________ M210912EN-C
The settings are temporary, and last only for a single session at a time.
On the next connection HMP155 reverts to the settings last set with the
SERI command.
SMODE
Use the SMODE command to set the user port to permanent STOP,
RUN, POLL or SEND mode.
SMODE [xxxx]
where
xxxx= STOP, RUN, POLL or SEND
Example:
>smode
Serial mode : STOP ? POLL
>
Table 12
Selection of Output Modes
Mode Output
Available
Commands
STOP
Only with the SEND command
All (default mode)
RUN
Automatic output
Only command S
POLL
Only with the SEND [addr] command
??, SEND [addr], OPEN [addr]
Selected output mode will be activated after power outages.
INTV
Use the INTV command to set the outputting interval for the RUN mode.
The default value for this command is 2 seconds.
INTV [xxx yyy]
where
xxx =
Output interval (0...255) where 0 produces an output whenever a
new measurement is made. Interval 1 s produces repeats of the
previous measurement if no newer measurement is available.
yyy =
Unit (s, min or h)
Example:
>intv 1 min
Interval : 1 min
>