Technical data, Specifications, Table 7 – Vaisala HMM212 User Manual
Page 33: Relative humidity specifications, Table 8, Temperature specifications, Table 9, Analog output specifications, Chapter 8

Chapter 8 ____________________________________________________________ Technical Data
CHAPTER 8
TECHNICAL DATA
This chapter provides the technical data of the product.
Specifications
Table 7
Relative Humidity Specifications
Property
Description / Value
Measurement range
0 ... 100 %RH
Accuracy (including non-linearity,
hysteresis and repeatability)
±2 %RH (0 … 90 %RH)
±3 %RH (90 … 100 %RH)
Response time (90%) at +20 °C in still air
(with sintered filter)
60 s
Typical temperature dependence of
electronics
0.02 %RH/°C
Sensor
HUMICAP
180R
Table 8
Temperature Specifications
Property
Description / Value
Measurement range
-70 ... +180 °C
Typical accuracy of electronics at +20 °C ±0.1 °C
Typical temperature dependence of
electronics
0.0025 °C/°C
Sensor
Pt 1000 RTD 1/3 Class B IEC 751
Table 9
Analog Output Specifications
Property
Description / Value
Two analog outputs
4 ... 20 mA
Ch1: RH
Ch2: T
Typical accuracy
±0.1 %FS
Resolution
Ch1: 0.025 %RH
Ch2: 0.01 %FS
Typical temperature dependence
0.005 %FS/°C
VAISALA _______________________________________________________________________ 33