Chapter 5 _______________________________________________________________ Calibration
Calibration Table
Table 5
Greenspan's Calibration Table with Output Values
According to the Chosen Scale
Temperature
LiCl
°C
°F
%RH
15
59
*)
20
68
11.3
25
77
11.3
30
86
11.3
35
95
11.3
4 ... 20 mA
mA
5.81
5.81
5.81
5.81
0 ... 20 mA
mA
2.26
2.26
2.26
2.26
0 ... 1 V
V
0.113
0.113
0.113
0.113
0 ... 5 V
V
0.565
0.565
0.565
0.565
0 ... 10 V
V
1.13
1.13
1.13
1.13
Temperature
NaCl
°C
°F
%RH
15
59
75.6
20
68
75.5
25
77
75.3
30
86
75.1
35
95
74.9
4 ... 20 mA
mA
16.10
16.08
16.05
16.02
15.98
0 ... 20 mA
mA
15.12
15.10
15.06
15.02
14.98
0 ... 1 V
V
0.756
0.755
0.753
0.751
0.749
0 ... 5 V
V
3.780
3.775
3.765
3.755
3.745
0 ... 10 V
V
7.56
7.55
7.53
7.51
7.49
*) If the LiCl solution is used or stored in temperatures below +18 °C (+64 °F), the equilibrium
humidity of the salt solution may change permanently.