Chapter 5 _______________________________________________________________ Calibration

Calibration Table

Table 5

Greenspan's Calibration Table with Output Values
According to the Chosen Scale

Temperature

LiCl

°C
°F
%RH

15
59
*)

20
68
11.3

25
77
11.3

30
86
11.3

35
95
11.3

4 ... 20 mA

mA

5.81

5.81

5.81

5.81

0 ... 20 mA

mA

2.26

2.26

2.26

2.26

0 ... 1 V

V

0.113

0.113

0.113

0.113

0 ... 5 V

V

0.565

0.565

0.565

0.565

0 ... 10 V

V

1.13

1.13

1.13

1.13

Temperature

NaCl

°C
°F
%RH

15
59
75.6

20
68
75.5

25
77
75.3

30
86
75.1

35
95
74.9

4 ... 20 mA

mA

16.10

16.08

16.05

16.02

15.98

0 ... 20 mA

mA

15.12

15.10

15.06

15.02

14.98

0 ... 1 V

V

0.756

0.755

0.753

0.751

0.749

0 ... 5 V

V

3.780

3.775

3.765

3.755

3.745

0 ... 10 V

V

7.56

7.55

7.53

7.51

7.49

*) If the LiCl solution is used or stored in temperatures below +18 °C (+64 °F), the equilibrium

humidity of the salt solution may change permanently.

VAISALA _______________________________________________________________________ 25