Chapter 7 _______________________________________________ Technical Data
VAISALA _________________________________________________________ 35
CHAPTER 7
TECHNICAL DATA
Operating temperature range
+0 ... +50 ºC
NOTE
If the LiCl solution is used or stored in temperatures below
+18 ºC, its equilibrium humidity changes permanently.
Accuracy:
lithium chloride LiCl
±1.0 %RH +
Greenspan's
uncertainty*
magnesium chloride MgCl
2
±1.0 %RH +
Greenspan's
uncertainty*
natrium chloride NaCl
±1.4 %RH +
Greenspan's
uncertainty*
potassium sulphate K
2
SO
4
±1.5 %RH +
Greenspan's
uncertainty*
*) The uncertainty given in Greenspan's calibration table (on page 25) at the calibration
temperature, for example the accuracy of LiCl salt at +20 °C is ±(1.0 + 0.3) %RH
= ±1.3 %RH
- Calibration of Digital Transmitters with HMI41 (36 pages)
- Calibration of Series HMDW2030 and HMP130 Transmitter with HMI41 (14 pages)
- Calibration of Series HMDW6070 and HMP140 Transmitter with HMI41 (30 pages)
- HM34 (30 pages)
- HM40 (47 pages)
- HM44 (52 pages)
- HM70 (83 pages)
- HMD40 (1 page)
- HMD60 (4 pages)
- HMD70 (18 pages)
- HMDW110 (62 pages)
- HMDW80 (51 pages)
- HMI41 (74 pages)
- HMP41 (72 pages)
- HMM100 (71 pages)
- HMM105 (23 pages)
- HMM211 (42 pages)
- HMM212 (36 pages)
- HMM213 (52 pages)
- HMP140 (28 pages)
- HMP155 (84 pages)
- HMP228 (115 pages)
- HMP230 (163 pages)
- HMP240 (130 pages)
- HMP260 (118 pages)
- HMP60 (71 pages)
- HMT100 (52 pages)
- HMT120 (87 pages)
- HMT130 (95 pages)
- HMT140 (76 pages)
- HMT310 (88 pages)
- HMT310 (105 pages)
- HMT330 (209 pages)
- HMT360 (97 pages)
- HMT360 (63 pages)
- HMT360N (110 pages)
- HMW40 (1 page)
- HMW90 (110 pages)
- SHM40 (68 pages)
- RDP100 (14 pages)