Chapter 7 _______________________________________________ Technical Data

VAISALA _________________________________________________________ 35

CHAPTER 7

TECHNICAL DATA

Operating temperature range

+0 ... +50 ºC

NOTE

If the LiCl solution is used or stored in temperatures below
+18 ºC, its equilibrium humidity changes permanently.


Accuracy:

lithium chloride LiCl

±1.0 %RH +
Greenspan's
uncertainty*

magnesium chloride MgCl

2

±1.0 %RH +
Greenspan's
uncertainty*

natrium chloride NaCl

±1.4 %RH +
Greenspan's
uncertainty*

potassium sulphate K

2

SO

4

±1.5 %RH +
Greenspan's
uncertainty*

*) The uncertainty given in Greenspan's calibration table (on page 25) at the calibration

temperature, for example the accuracy of LiCl salt at +20 °C is ±(1.0 + 0.3) %RH
= ±1.3 %RH