Sper Scientific 300055 Digital Refractometer Automotive User Manual
Page 6
- 6 -
•
Allow the refractometer and the distilled
water to acclimate to the ambient
environment.
•
Clean the PRISM (3) before and after
every use. There should be no spots or
film on the prism.
•
Drop at least 1.0 ml of distilled water (0.0)
onto the PRISM (3).
•
Close the COVER (2).
•
Press the READ (9) button to turn on the
refractometer.
•
Press the ZERO (10) button for 3 seconds
to enter calibration mode.
•
"CAL" flashes on the display. There will
be a red flash in the PRISM (3).
•
The calibration is complete after "End" is
displayed.
5. MEASUREMENT
•
Clean the SAMPLE PLATE (1) and
PRISM
(3) before and after every use.
There should be no film or spots on the
prism. If residue remains from a previous
test, it will affect the accuracy of the unit.
•
When the temperature difference between
the ambient temperature and the
refractometer is greater than 10°F (5°C),
- 7 -
place the refractometer in the lab and
allow it to adjust to the ambient
temperature. Allow apx. 30 minutes for
every 5°F (10°C)
difference in temperature.
•
Press the READ (9) button to turn the unit
on and display the temperature. This unit
features Automatic Temperature Compen-
sation (ATC) within the range of 41~104°F
(5~40°C), eliminating the need for manual
temperature correction.
•
Perform a calibration (pg 6) before the
day's first use or after extended daily use.
•
Fill the PRISM (3) with about 1.0ml of the
sample solution.
•
Close the COVER (2).
•
Press the READ (9) button immediately to
obtain a reading. Delays may cause the
sample to evaporate.
•
Within 1 second, the value is displayed.
•
When the reading is under-range, the low
indicator (LLL) is displayed. The high
indicator (HHH) is displayed when the
reading is over-range.
Models with Multiple Scales
•
Follow the measurement instructions