Sper Scientific 300055 Digital Refractometer Automotive User Manual

Page 6

background image

- 6 -

Allow the refractometer and the distilled
water to acclimate to the ambient
environment.

Clean the PRISM (3) before and after
every use. There should be no spots or
film on the prism.

Drop at least 1.0 ml of distilled water (0.0)
onto the PRISM (3).

Close the COVER (2).

Press the READ (9) button to turn on the
refractometer.

Press the ZERO (10) button for 3 seconds
to enter calibration mode.

"CAL" flashes on the display. There will
be a red flash in the PRISM (3).

The calibration is complete after "End" is
displayed.

5. MEASUREMENT

Clean the SAMPLE PLATE (1) and
PRISM

(3) before and after every use.

There should be no film or spots on the
prism. If residue remains from a previous
test, it will affect the accuracy of the unit.

When the temperature difference between
the ambient temperature and the
refractometer is greater than 10°F (5°C),

- 7 -

place the refractometer in the lab and
allow it to adjust to the ambient
temperature. Allow apx. 30 minutes for
every 5°F (10°C)
difference in temperature.

Press the READ (9) button to turn the unit
on and display the temperature. This unit
features Automatic Temperature Compen-
sation (ATC) within the range of 41~104°F
(5~40°C), eliminating the need for manual
temperature correction.

Perform a calibration (pg 6) before the
day's first use or after extended daily use.

Fill the PRISM (3) with about 1.0ml of the
sample solution.

Close the COVER (2).

Press the READ (9) button immediately to
obtain a reading. Delays may cause the
sample to evaporate.

Within 1 second, the value is displayed.

When the reading is under-range, the low
indicator (LLL) is displayed. The high
indicator (HHH) is displayed when the
reading is over-range.


Models with Multiple Scales

Follow the measurement instructions