4. OPERATING INSTRUCTIONS
A. General Measurement
To avoid damage, it is important to clean the
PRISM WINDOW
before and after each use.
Press the ON/OFF button to turn the
Refractometer ON. The unit will automatically
shut off after about 3 minutes of inactivity.
Perform the Calibration Procedure (4B)
With the unit on, fill the PRISM WINDOW with
the sample to be measured and then close the
COVER
.
Press the READ button immediately to ensure
accurate readings. Delays may cause the sample
to evaporate.
B. Calibration Procedure
Clean the PRISM WINDOW before and after
each use.
Fill the PRISM WINDOW with a standard solution
such as distilled or deionized water. The ideal
temperature of the solution and the unit should
be 65°~72°F (18°~22°C).
Close the COVER.
Press the ON/OFF button to turn the unit ON.
Press the ZERO button for 3 seconds.
When the procedure is complete, CAL will be
displayed.
To check the nD value, press the ZERO button
once.
This data can be viewed at any time to show the
last calibration value.
A flashing crosshair symbol is displayed when
the calibration solution falls outside the accepted
range.
C. Zero Checking
Fill the prism window with distilled water solution
that is at the same temperature as the solution to
be tested.
Close the COVER