Sper Scientific 300037 Refractometer - Programmable User Manual

Page 4

background image

4

4. OPERATING INSTRUCTIONS

A. General Measurement



To avoid damage, it is important to clean the
PRISM WINDOW

before and after each use.



Press the ON/OFF button to turn the
Refractometer ON. The unit will automatically
shut off after about 3 minutes of inactivity.



Perform the Calibration Procedure (4B)



With the unit on, fill the PRISM WINDOW with
the sample to be measured and then close the
COVER

.



Press the READ button immediately to ensure
accurate readings. Delays may cause the sample
to evaporate.

B. Calibration Procedure



Clean the PRISM WINDOW before and after
each use.



Fill the PRISM WINDOW with a standard solution
such as distilled or deionized water. The ideal
temperature of the solution and the unit should
be 65°~72°F (18°~22°C).



Close the COVER.



Press the ON/OFF button to turn the unit ON.



Press the ZERO button for 3 seconds.



When the procedure is complete, CAL will be
displayed.



To check the nD value, press the ZERO button
once.



This data can be viewed at any time to show the
last calibration value.



A flashing crosshair symbol is displayed when
the calibration solution falls outside the accepted
range.

C. Zero Checking



Fill the prism window with distilled water solution
that is at the same temperature as the solution to
be tested.



Close the COVER