LARGE DISPLAY

ATC PH PENS

850051 and 850052










































WARRANTY: Sper Scientific warrants
this product against defects in materials
a n d w o r k m a n s h i p f o r a
period of 1 year from the date of
purchase, and agrees to repair or
replace any defective unit without
charge. If your model has since been
discontinued, an equivalent Sper
Scientific product will be substituted if
available. This warranty does not cover
probes, batteries, or damage resulting
from accident, misuse, or abuse. In or-
der to obtain warranty service, ship the
unit postage prepaid to:

SPER SCIENTIFIC

7720 E Redfield Rd, Ste 7

Scottsdale, AZ 85260

Email: [email protected]

Include a description of the problem and
your return address along with the re-
t u r n e d i t e m . R e t u r n t h e
warranty card within 10 days or
register at: www.sperscientific.com.

BATTERY REPLACEMENT

• Replace all 4 LR44 batteries when the

low battery icon is flashing, or when
troubleshooting (below) fails.

• Unscrew the cover and install the

batteries with the flat positive (+) side
of the batteries facing the spring.

• Do not remove the O-ring.

• Remove the batteries when the unit is

not in use for an extended period.

TROUBLESHOOTING

• If there is no display, or you have

erratic readings after having soaked
the electrode in distilled/de-ionized
water for at least 1 hour prior to
calibration, check the continuity of the
batteries by removing them for 1
minute, then reinsert.

• Slow Response: Soak the electrode in

tap water for 10-15 minutes, then rinse
with distilled/de-ionized water.

• "- -

-

" displays when the pH sample is

too acidic or too alkaline.

• "H" or "L" displays when the

temperature is too hot (H) or too cold
(L).

• No Display: Be sure to press for at

least one second. If the problem
persists, see Battery Replacement
(above).

• If your readings are still erratic after

calibration, soak the electrode for 10
to 30 minutes in distilled water.

• Note: It is normal for the pH value to

fluctuate quickly when the electrode is
exposed to air and not immersed.

SPECIFICATIONS

pH Range

0~14 pH

pH Resolution
850051
850052


0.1 pH
0.01 pH


0.5 °F/°C
0.1°F/°C

Accuracy
850051
850052


±0.2 pH
±0.05 pH


±1°C
±0.5°C

Adjustable
Calibration
Points

pH 4
pH 7
pH 10

3.5~4.5
6.5~7.5
9.5~10.5

Temp.
Range
Operating
Environ.

31~122°F
0~50°C

<80 RH

Weight

2.4 oz, 68g with battery

Dimension

6½" x 1½" x 1"
165 x 35 x 32 mm

0~60°C