LARGE DISPLAY PH PEN

850050









































WARRANTY:

Sper Scientific warrants

this product against defects in materials
and workmanship for a period of 1 year
from the date of purchase, and agrees
to repair or replace any defective unit
without charge. If your model has since
been discontinued, an equivalent Sper
Scientific product will be substituted if
available. This warranty does not cover
probes, batteries, or damage resulting
from accident, misuse, or abuse. To
obtain warranty service, ship the unit
postage prepaid to:

SPER SCIENTIFIC

7720 E Redfield Rd, Ste 7

Scottsdale, AZ 85260

Email: [email protected]

Include a description of the problem
and your return address along with the
returned item. Return the warranty card
within 10 days or register at:

www.sperscientific.com

BATTERY REPLACEMENT

 Replace all 4 LR44 batteries when the

low battery icon is flashing, or when
troubleshooting (below) fails.

 Unscrew the cover and install the

batteries with the flat positive (+) side
of the batteries facing the spring.

 Do not remove the O-ring.

 Remove the batteries when the unit is

not in use for an extended period.

TROUBLESHOOTING

 If there is no display, or you have

erratic readings after having soaked
the electrode in distilled/de-ionized
water for at least 1 hour prior to
calibration, check the continuity of the
batteries by removing them for 1
minute, then reinsert.

 Slow Response: Soak the electrode in

tap water for 10-15 minutes, then rinse
with distilled/de-ionized water.

 "- -

-

" displays when the pH sample is

too acidic or too alkaline.

 No Display: Be sure to press for at

least one second. If the problem
persists, see Battery Replacement
(above).

 If your readings are still erratic after

calibration, soak the electrode for 10
to 30 minutes in distilled water.

 Note: It is normal for the pH value to

fluctuate quickly when the electrode is
exposed to air and not immersed.

SPECIFICATIONS

Range

2~12 pH

Resolution

0.1 pH

Accuracy

Adjustable
Calibration
Points

pH 4
pH 7
pH 10

3.5~4.5
6.5~7.5
9.5~10.5

Operating
Environ.

31~122°F
0~50°C

<80 RH

Weight

2.4 oz, 68g with battery

Dimension

6½" x 1½" x 1"
165 x 35 x 32 mm

±0.3 pH