- 6 -

3-H. RS232 PC SERIAL INTERFACE

The unit features an RS232 output via 3.5 mm Terminal. The output is a

16 digit data stream which can be utilized by user's specific application. An
RS232 lead with the following connection will be required to link the
insrument with the PC serial input.

Meter (3.5 mm jack plug)

PC (9W "D" Connector)

Center Pin

Pin 2

Ground/Shield

Pin 5

The 16 digit data stream will be displayed in the following format.

D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1

DO

End Word

D1 to D8

Display reading, D1=LSD, D8=MSD

D9

Decimal Point(DP), position from right to the left

0 = No DP, 1 = 1 DP, 2 = 2 DP, 3 = 3 DP

D10

Polarity: 0 = Positive, 1 = Negative

Annunciator for Display

Psi = 23

mbar = 86

uHg = 77

mm/Hg = 78

Pa = 87

torr = 90

inch/Hg = 25

K Pa = 88

D13

1

D14

4

D15

Start Word

RS232 FORMAT: 9600, N, 8, 1

D11 & D12