Rh calibration – Sper Scientific 800046 Indoor Air Quality Meter User Manual

Page 14

background image

- 14 -

The meter defaults to calibration levels of 33% humidity and
75% salt solution. The recommended ambient condition is
25°C with stable humidity (close to the calibrating value). To
abort RH calibration, turn the meter off.

Caution…
Do not calibrate humidity without the calibration salt as
doing so may cause permanent damage. Contact Sper
Scientific for calibration salt or services.

RH CALIBRATION

33% Calibration

1. Plug the sensor probe into a 33% salt bottle.

2. With the meter in Normal mode, hold CAL and DP/WBT

simultaneously to enter the 33% calibration mode.


3. “CAL” and the calibrating value (32.7% if at 25

°

C) will

blink on the LCD with the current temperature displayed
on the left.

4. Meter calibration is completed in approximately 60

minutes when “CAL” and humidity stop blinking, and the
meter returns to Normal mode.