Sound calibration – Sper Scientific 850069 Environmental Quality Meter w/Sound User Manual

Page 11

background image

- 11 -

This procedure requires an optional 2-
Point Acoustical Calibrator (850016):

1. Using the directions that accompany

the Acoustical Calibrator, ensure that
the calibrator is operational and is
equipped with the ½” microphone
adapter.

2. Insert meter’s sound probe into the

calibrator.

3. Turn the calibrator on and select 94

dB.

4. Turn the meter on and use the

FUNCTION

button to select the Sound

–Level meter function.

5. Press HOLD and MAX/MIN

simultaneously for 3 seconds and
release.

SOUND CALIBRATION