PROBE MAINTENANCE

Initial Use

1. Unpack the DO meter and fill the probe head with the

Electrolyte (OXEL-03) fluid provided.

2. Follow the steps shown in Figure 2.

Probe Head & Fluid Maintenance

1. Whenever the DO meter will not calibrate correctly or

the DO reading is unstable, check to see if the
Electrolyte fluid is dirty or has evaporated.

2. If the Electrolyte fluid needs to be replaced, refill the

probe head (completely), following the steps shown in
Figure 2 and recalibrate the DO meter (see page 3)

3. If the fluid has been replaced and the DO meter still

does not calibrate correctly or give a stable reading,
check the probe head diaphragm for damage.

4. The Teflon diaphragm is delicate and can be damaged if

it comes in contact with solid objects or if the meter is
dropped.