7-3 Calibration Test Using an External Weight

This function tests the balance weighing accuracy using an external weight and outputs the result.

This is available only when the “GLP output (info)” parameter is set to “1” or “2”. (Calibration test

does not perform calibration.)

Operation

1 Connect the AC adapter and warm up the balance

for at least 30 minutes with nothing on the pan.

2 Press and hold the CAL key until CCout is

displayed, and then release the key.

3 The

balance

displays CC 0 .

If you want to change the target weight (a list of

usable weights is shown on page 23), press the

SAMPLE key and proceed to step 4.

If you use the target weight value stored in the

balance, proceed to step 5.

4 Specify the target weight value as follows:

SAMPLE

key

To switch the display

condition to: “All of the

segments blinking” (target

weight selection mode) or

“The last two digits blinking”

(value adjustment mode).

RE-ZERO key (To increase the value)

MODE key (To decrease the value)

To select the target weight or

adjust the value.

PRINT

key

To store the new weight

value. Even if the AC

adapter is removed, the data

is maintained in non-volatile

memory.

CAL key

To cancel the operation and

return to CC 0 .