Meter problem solving – UEi Test Instruments SMART BELL PLUS KIT User Manual
Page 14

METER PROBLEM SOLVING
If any problems are not solved with these solutions, contact us or an authorized repair
center.
Fault symptom
Causes / Solutions
CO or CO2 too low
• Air leaking into probe, tubing, water trap
or connectors.
CO2 reading (- - - -)
• Meter was stored in a cold environment and
is not at normal working temperature.
Batteries not holding charge
• Batteries exhausted.
Meter does not respond to flue gas
• Particle filter blocked.
• Probe or tubing blocked.
• Pump not working or damaged with
contaminants.
Net temperature or Efficiency
• Ambient temperature set wrong during
calculation incorrect.
Automatic Calibration.
Flue temperature readings erratic
• Temperature plug reversed in socket.
• Faulty connection or break in cable or plug.
T flue displays (- - - -)
• Probe not connected or faulty
X-Air, EFF display
• CO2 reading is too low.
(- - - -)
12